Analysis: Optical
General Description:
The Cary 5000 combines PbSmart technology with the unparalleled optical design and performance of all Cary UV-Vis-NIR instruments. It requires only one detector to extend that performance into the NIR. With innovative PbSmart™ technology, the Cary 5000 extends the unparalleled Cary performance into the NIR to 3300 nm.
Technical Specification:
Monochromator : Double out of plane, littrow monochromator
Grating : Dual sided, 70x50mn, UV-VIS : 1200 lines/mm blazed at 250 nm, NIR: 300 lines/mm blazed at 1192 nm
Beam Splitting System : Chopper 30 Hz
Detectors : R928 PMT cooled Pbs Photocell
Limiting Resolution (nm):
UV-Vis <0.05
NIR <0.2
Stray Light (%T)
At 220 nm (10g/L Nal ASTM method) : <0.0008%
At 370 nm (50 mg/L NaNO2) : <0.0008%
At 200 nm(12 g/L KCl, TGA & BPEP method) : <1%
At 370 nm (0.25 g/L K2 Cr2 O7, TGA method) : <1%
At 1420 nm (H2O, 1 cm pathlength) : <0.0002%
At 2365 nm (CHCl3, 1 cm pathlength) : <0.00045%
Wavelength Range :
(N2 purge required below 187 nm) : 175-3300
Wavelength Reproducibility (nm):
Peak separation of repetitive scanning of a UV-Vis line source : <0.025
Peak separation of repetitive scanning of a NIR line source : <0.1
Standard deviation of 10 measurements, UV-Vis : <0.005
Standard deviation of 10 measurements, NIR : <0.02
Wavelength Accuracy (nm) :
UV-Vis : +/- 0.1
NIR : +/- 0.4
Photometric Accuracy (Abs):
Using double aperture at 0.3 Abs
UV-Vis : 0.0003
Photometric Linearity (Abs) All tests performed by addition of filters technique (UV-Vis:
500 nm, 1 sec Signal Averaging Time; NIR: 1200 nm, energy level 3)
UV-Vis, At 1 Abs : 0.0005
UV-Vis, At 2 Abs : 0.001
UV-Vis, At 3 Abs : 0.005
NIR, At 1 Abs : 0.001
NIR, At 2 Abs 0.005
Photometric Range (Abs) :
with RBA : 8
Photometric Display:
Abs : 9.9999
%T : 200.0000
Photometric Linearity (Abs) All tests performed by addition of filters technique (UV-Vis: 500 nm, 1 sec Signal Averaging Time ;
NIR: 1200 nm, energy level 3)
UV-Vis, At 1 Abs : 0.0005
UV-Vis, At 2 Abs : 0.001
UV-Vis, At 3 Abs : 0.005
NIR, At 1 Abs : 0.0005
NIR, At 2 Abs : 0.005
Photometric Range (Abs):
with RBA : 8
Photometric Display:
Abs : 9.9999
%T : 200.0000
Photometric Reproducibility (Abs) Using NIST 930D filters, at 590 nm, 2 nm SBW, 2 second Signal Averaging Time:
Maximum deviation at 1 Abs <0.0008
Standard deviation for 10 measurements : <0.00016
Using NIST 930D filters, at 546.1 nm, 2 nm SBW, 2 second Signal
Averaging Time:
Maximum deviation at 0.5 Abs : <0.0004
Standard deviation of 10 measurements : <0.00008
Using NIST 930D filters, at 546.1 nm, 2 nm SBW, 2 second Signal
Averaging Time:
Maximum deviation at 0.5 Abs : <0.0004
Standard deviation of 10 measurements : <0.00008
Reference: Manual book for UV-VIS-NIR cary 4000,5000,6000i Spectrophotometers.