Analysis: Optical

UV-Visible Spectrophotometer Model Cary 5000 UV-VIS-NIR

36. UV Visible Spectrophotometer Model Cary 5000 UV NIS NIRGeneral Description:

The Cary 5000 combines PbSmart technology with the unparalleled optical design and performance of all Cary UV-Vis-NIR instruments. It requires only one detector to extend that performance into the NIR. With innovative PbSmart™ technology, the Cary 5000 extends the unparalleled Cary performance into the NIR to 3300 nm.

Technical Specification:

   Monochromator : Double out of plane, littrow monochromator

   Grating : Dual sided, 70x50mn, UV-VIS : 1200 lines/mm blazed at 250 nm, NIR: 300 lines/mm blazed at 1192 nm

   Beam Splitting System : Chopper 30 Hz

   Detectors : R928 PMT cooled Pbs Photocell

   Limiting Resolution (nm):

   UV-Vis <0.05

   NIR <0.2

   Stray Light (%T)

   At 220 nm (10g/L Nal ASTM method) : <0.0008%

   At 370 nm (50 mg/L NaNO2) : <0.0008%

   At 200 nm(12 g/L KCl, TGA & BPEP method) : <1%

   At 370 nm (0.25 g/L K2 Cr2 O7, TGA method) : <1%

   At 1420 nm (H2O, 1 cm pathlength) : <0.0002%

   At 2365 nm (CHCl3, 1 cm pathlength) : <0.00045%

   Wavelength Range :

   (N2 purge required below 187 nm) : 175-3300

   Wavelength Reproducibility (nm):

   Peak separation of repetitive scanning of a UV-Vis line source : <0.025

   Peak separation of repetitive scanning of a NIR line source : <0.1

   Standard deviation of 10 measurements, UV-Vis : <0.005

   Standard deviation of 10 measurements, NIR : <0.02

   Wavelength Accuracy (nm) :

       UV-Vis : +/- 0.1

       NIR : +/- 0.4

   Photometric Accuracy (Abs):

       Using double aperture at 0.3 Abs

       UV-Vis : 0.0003

   Photometric Linearity (Abs) All tests performed by addition of filters technique (UV-Vis:

       500 nm, 1 sec Signal Averaging Time; NIR: 1200 nm, energy level 3)

       UV-Vis, At 1 Abs : 0.0005

       UV-Vis, At 2 Abs : 0.001

       UV-Vis, At 3 Abs : 0.005

       NIR, At 1 Abs : 0.001

       NIR, At 2 Abs 0.005

       Photometric Range (Abs) :

           with RBA : 8

   Photometric Display:

       Abs : 9.9999

       %T : 200.0000

   Photometric Linearity (Abs) All tests performed by addition of filters technique (UV-Vis: 500 nm, 1 sec Signal Averaging Time ;

       NIR: 1200 nm, energy level 3)

       UV-Vis, At 1 Abs : 0.0005

       UV-Vis, At 2 Abs : 0.001

       UV-Vis, At 3 Abs : 0.005

       NIR, At 1 Abs : 0.0005

       NIR, At 2 Abs : 0.005

   Photometric Range (Abs):

       with RBA : 8

   Photometric Display:

       Abs : 9.9999

       %T : 200.0000

   Photometric Reproducibility (Abs) Using NIST 930D filters, at 590 nm, 2 nm SBW, 2 second Signal Averaging Time:

       Maximum deviation at 1 Abs <0.0008

       Standard deviation for 10 measurements : <0.00016

       Using NIST 930D filters, at 546.1 nm, 2 nm SBW, 2 second Signal

       Averaging Time:

       Maximum deviation at 0.5 Abs : <0.0004

       Standard deviation of 10 measurements : <0.00008

       Using NIST 930D filters, at 546.1 nm, 2 nm SBW, 2 second Signal

       Averaging Time:

       Maximum deviation at 0.5 Abs : <0.0004

       Standard deviation of 10 measurements : <0.00008

Reference: Manual book for UV-VIS-NIR cary 4000,5000,6000i Spectrophotometers.